Giriş çalışma gerilimi 90-264Vac (100-240Vac nominal) able to withstand input voltages of up to 300Vac for up to 5 seconds |
|
BF ready medical isolation (MOPP) |
|
Class I / II with level B EMC |
|
1U profile |
|
Low airflow requirement |
|
BF ready medical isolation (MOPP) |
|
Efficiency up to 92% At 230Vac, 100% load |
|
Short circuit protection Hiccup mode, auto recovery after removal of short circuit |
|
Over Temperature protection Latching, need to cycle ac to restart unit |
|
Over Voltage Protection Latching, need to cycle ac to restart unit. |
|
Remote on/of |
|
Temperature 0°C to 50°C operational, Low Temp Startup -20°C |
|
Shock Conforms to MIL-STD-810G, Method 516.6, Pro IV, VI |
|
Vibration Conforms to MIL-STD-810G, Method 514.4, Pro I, Cat 1,9 |
|
Altitude -200 to 5000 metres operational (3000m for 60601-1) (-200 to 5000m storage/transportation) |
|
|